Testing of actual/suspected device: (10/213): a getinge field service engineer (fse) evaluated the iabp unit, upon arrival he found that the iabp had no line power and was not charging the batteries, he found that the line cord reel had separated and was not passing the 120 volts to the power supply.
To fix the issue, the fse replaced the reel, ac power cord and the cardiosave sr.
Power supply assy, then performed a full pm with calibration, functional and safety checks to meet factory specifications.
The iabp was then released to the customer and cleared for clinical service.
A supplemental report will be submitted upon completion of our investigation.