• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC HEARTSTART MRX MONITOR/DEFIB; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC HEARTSTART MRX MONITOR/DEFIB; DEFIBRILLATOR Back to Search Results
Model Number M3535A
Device Problems Failure of Device to Self-Test (2937); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/08/2021
Event Type  malfunction  
Event Description
It was reported to philips that the device required a technical check after the functional test.There was no patient involvement.
 
Manufacturer Narrative
The customer requested that a philips field service engineer (fse) be dispatched to the customer site.The reported issue was confirmed and traced nibp calibration being overdue.Upon conclusion of the evaluation, it was determined that the nibp calibration was overdue which is not a malfunction of the device but rather due to a lack of maintenance.The fse calibrated the nibp to resolve the reported issue.The device remains at the customer site and no further evaluation is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX MONITOR/DEFIB
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
daniel derochers
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key12264476
MDR Text Key264699755
Report Number3030677-2021-13345
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838000018
UDI-Public00884838000018
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3535A
Device Catalogue NumberM3535A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date07/08/2021
Date Manufacturer Received07/08/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/21/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-