• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 X-PORT ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 X-PORT ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 0607540CE
Device Problems Unsealed Device Packaging (1444); Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/05/2021
Event Type  malfunction  
Manufacturer Narrative
The catalog number has not been cleared in the us but is similar to the x-port isp implantable port, chronoflex single-lumen, 8f that are cleared in the us.The pro code and 510 k number for the x-port isp implantable port, chronoflex single-lumen, 8f are identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, photos were provided for review.The investigation of the reported event is currently underway.(expiry date: 01/2023).
 
Event Description
It was reported that prior to port placement, double-wrapped box was empty.It was further reported that the packaging allegedly had no seal.There was no patient contact.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one unsealed x-port mri implantable port kit was returned for evaluation and three electronic photos were provided for review.Visual evaluation was performed.The inner package of the sample arrived unsealed.Ifus, id cards, implant record cards and product specifications were received within package.No components were available inside the inner package.However, the investigation is inconclusive for the reported component missing and unsealed device package issues as the package was returned unsealed and the exact circumstances at the time of the reported event are unknown and the provided photos are not sufficient enough to confirm the issue.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the x-port isp implantable port, chronoflex single-lumen, 8f that are cleared in the us.The pro code and 510 k number for the x-port isp implantable port, chronoflex single-lumen, 8f are identified in d2 and g4.H10: d4 (expiry date: 01/2023).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event Description
It was reported that prior to port placement, double-wrapped box was empty.It was further reported that the packaging allegedly had no seal.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-PORT ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12265186
MDR Text Key264695708
Report Number3006260740-2021-03083
Device Sequence Number1
Product Code LJT
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K022983
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0607540CE
Device Lot NumberREFP2328
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-