It was reported that iag bc pro global pnk 20ga x 1.
0in was not able to connect.
The following information was provided by the initial reporter: on the (b)(6) 2021, in the hospital carried out a material vigilance on the 20g catheter.
"catheter set up for an injected scan.
When connecting the catheter to the infusion tubing, impossible to connect the catheter to the tubing.
After testing, no problem with the tubing, it is the luer tip of the catheter that causes the problem.
Need to change the catheter and therefore put the patient back on.
|