Model Number 550AHCT
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2021
Event Type  malfunction  
Manufacturer Narrative
Per the manufacturer's sales associate, the end user had been doing in-vivo calibrations and it took several times before the unit values were closer to the blood gas values.
 
Event Description
It was reported that during use of the device for cardiopulmonary bypass (cpb), the blood parameter monitor (bpm) was displaying a different hematocrit (hct) value than the blood gas analyzer.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
Updated block: b5.
 
Event Description
Per clinical review: the team had an incident with the blood parameter monitor (bpm), during a cardiopulmonary bypass (cpb) procedure on (b)(6) 2021.The bpm passed the color chip test without issue, for the procedure on that day.After the first in-vivo, the team stated, that the hematocrit (hct) values or numbers were way off from the blood gas values.The in-vivo calibrations were being done.And took several times before the unit trended closer to blood gases.The team did not exchange the unit out.There was no delay in the continuation of the surgical procedure.
 
Manufacturer Narrative
Updated blocks: h3 and h6 the reported complaint was not verifiable since the product was not returned for evaluation or testing.Multiple diligence attempts for part return were unsuccessful.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Manufacturer Narrative
Updated blocks: h3 and h6 during laboratory analysis, the product surveillance technician (pst) performed numerous in-vivo adjustments and each one was reflected accurately on the display.The monitor was turned off and set idle for eight hours, and restarted in operate mode for an extended period of time with no issues.There was no way to recreate the specific situation reported in the complaint but the issue was not duplicated.
 
Manufacturer Narrative
Updated blocks: d9 and h3 h3 81 - evaluation is in progress, but not yet concluded.
 
Manufacturer Narrative
The reported complaint could not be confirmed.The service repair technician (srt) was not able to duplicate the reported complaint.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 550
Type of Device
CARDIOPULMONARY BYPASS ON-LINE BLOOD GAS
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key12273931
MDR Text Key264951388
Report Number1828100-2021-00262
Device Sequence Number1
Product Code DRY
UDI-Device Identifier00886799001790
UDI-Public(01)00886799001790(11)201118
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number550AHCT
Device Catalogue Number550AHCT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/18/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
BLOOD GAS ANALYZER; BLOOD GAS ANALYZER.
-
-