MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/13/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: a getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue, to fix the issue the fse replaced the cart power cord and power supply. After replacing he confirmed that both carts charged up the unit and the ac led light was on. He then performed all functional and safety checks to meet factory specifications. Unit passed all functional and safety test per factory specifications. The iabp was then released to the customer and cleared for clinical service. A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) was not charging. It is unknown the circumstances under which the event occurred. It is also unknown if there was a patient involvement. However, there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE HYBRID TYPE B PLUG
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12274356
MDR Text Key264970724
Report Number2249723-2021-01712
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 11/19/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/28/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse

-
-