DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID DOMESTIC SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/12/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device (10/213): the getinge field service engineer (fse) that encountered the issue replaced the power management board and completed the pm with full calibration, functional and safety checks to meet factory specifications.
Unit passed all calibration, functional and safety test per factory specifications.
The iabp was then released to the customer and cleared for clinical service.
A supplemental report will be submitted if this information is provided to us.
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the cardiosave intra-aortic balloon pump (iabp) would not charge the batteries.
There was no patient involvement, and no adverse event reported.
|
|