• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. DISP FIRSTPASS STR PASSR SELF; PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. DISP FIRSTPASS STR PASSR SELF; PASSER Back to Search Results
Catalog Number 22-4038
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/14/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during use in an unknown procedure, the firstpass suture passer broke upon use in surgery.No pieces fell into the patient.The procedure was successfully completed without delay using a smith and nephew back up device.No patient injury or other complications were reported.No more information is available.
 
Manufacturer Narrative
H10 h6 the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.A review of the customer provided image shows a device label confirming part number and lot number.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISP FIRSTPASS STR PASSR SELF
Type of Device
PASSER
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key12291408
MDR Text Key265575503
Report Number3006524618-2021-00811
Device Sequence Number1
Product Code HWQ
UDI-Device Identifier00885556724545
UDI-Public885556724545
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 09/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/03/2023
Device Catalogue Number22-4038
Device Lot Number2065225
Was Device Available for Evaluation? No
Date Manufacturer Received09/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-