This is filed to report difficult to remove, delay, prolonged hospitalization, and surgical intervention.
It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.
A steerable guide catheter (sgc) was advanced to the mitral valve without issue.
Then a clip delivery system (cds) was inserted in the sgc and advanced to the mitral valve.
The clip grasped the leaflets fine; however, the clip would not completely close.
It was noted that the arm positioner felt extremely tight and the clip would only close half way.
Troubleshooting was performed and the clip appeared to be functioning properly.
Therefore, the clip re-grasped the leaflets and the clip was locked sooner this time.
However, the arm positioner felt tight again and the clip would not fully close.
The clip was inverted and retracted back to the left atrium.
The clip would still not fully close and so the clip was forced closed.
The clip seemed closed, but when the clip was retracted back into the guide, the mandrel shaft separated and the clip became stuck at the tip of the guide.
The soft tip of the sgc looked asymmetrical.
Both the sgc and cds were being removed as a single unit, but the clip became stuck at the femoral groin causing a clinically significant delay.
Tissue was observed in the clip.
The physician stated that the tissue possibly came from the femoral vein.
The patient remained hospitalized and a surgical cut-down was performed to remove the sgc and cds.
The patient is stable and is pending another mitraclip procedure.
No clips were implanted, and mr is 4.
There was no clinically significant delay in the procedure.
No additional information was provided.
|