It was reported to philips that the device has a fault during discharge, resistance load value low.The field service engineer (fse) evaluated the device and confirmed the discharge malfunction.The device needs a high voltage capacitor.Resolution: the intervention started only after confirmation of the disinfection and/or decontamination by the health operator.Customer reports the following fault: checked by clinical engineering, it discharges 43ohm out of 50.Replaced condenser functional tests carried out with positive outcome.The device was left operational regarding reported defect.Upon conclusion of the evaluation, it was determined that the high voltage capacitor.It was replaced.The device passed testing and was put back into service.
|