Model Number 0684-00-0576-01 |
Device Problems
Inappropriate Waveform (2536); Pressure Problem (3012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/20/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
|
Event Description
|
It was reported that after insertion of the intra-aortic balloon (iab), the customer noticed the arterial line waveform was very dampened, no pressure readings were displayed, and the console was generating an unable to calibrate optical sensor alarm.The customer attempted to reconnect and attempted to invoke a calibration but the issue persisted.A flush line and transducer were connected to the catheter so the customer was advised to disconnect the fiber optic sensor and zero the transducer.A waveform was present, the pump zeroed, and they resumed pumping alarm free.The fiber optic remained disconnected.There was no patient harm or adverse event reported.
|
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If additional information is provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period aug-2019 through jul-2021 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #(b)(4).
|
|