On 08/09/2021, we received medwatch report # mw5102641 from the user facility.We reviewed our records and confirmed the report was related to the subject event.On (b)(6) 2021, a steris service technician arrived onsite to inspect the table and found evidence of fluid intrusion within the power supply.Upon further inspection, the technician also found that the table's shrouds were damaged.The damage to the shrouds allowed fluids from the ongoing procedure to enter the table base and contact the power supply causing a short resulting in the reported event.Based on the technician's inspection, the damage observed to the table's shrouds is consistent with items being stored on the base of the table.The 4085 surgical table operator manual states (1-3)," do not store items on table base.Doing so may result in equipment damage or inadvertent tabletop movement placing the patient and/or user at risk of injury." the technician made the necessary repairs, tested the unit, confirmed it was operating according to specification, and returned to service.A steris account manager performed an in-service training on (b)(6) 2021 on the proper maintenance and operation of the 4085 surgical table, specifically to not store items on the table base.No additional issues have been reported.
|