SMITH & NEPHEW, INC. UNIVERSAL PIN DRIVER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 71513331
Device Problems Material Fragmentation (1261); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/27/2021
Event Type  malfunction  
Event Description
It was reported that, during a tka surgery, two (2) universal pin drivers failed to hold the profix bone spikes, both cracked while the surgeon placed the pins into the a/p block.Pieces of the devices were retrieved but it is unknown how.Surgery was resumed, without any delay, with a back-up device.Patient was not injured as consequence of this problem.Also, a genesis ii posterior stabilized constrained housing reamer dome have become dull and did not ream well.Surgery was resumed, without any delay, with the same device.Patient was not harmed as consequence of this problem.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL PIN DRIVER
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12307366
MDR Text Key266015461
Report Number1020279-2021-06289
Device Sequence Number1
Product Code LXH
UDI-Device Identifier03596010589064
UDI-Public03596010589064
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71513331
Device Catalogue Number71513331
Device Lot Number18BM18534
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-