Model Number TPL0059
Device Problem Imprecision (1307)
Patient Problem Unspecified Nervous System Problem (4426)
Event Date 07/19/2021
Event Type  Injury  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment. Testing revealed that the guidance system passed a stress test, built in self test, and advanced accuracy test. No failures were found. The system then passed the system checkout and was found to be fully functional. If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a guidance system being used during a spinal procedure. It was reported that two weeks after an l3-l5 procedure, the patient experienced foot drop symptoms. After imaging, the right l4 screw was found to be medial by 4-5 mm. The manufacturer representative stated that the patient had high bmi and there was soft tissue pressure on the tools during the procedure. There was no skive potential in the planning for the trajectories. A revision procedure to reposition the screw using navigation was planned. There was no delay to the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMAZOR X SYSTEM
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il 30795 67
IS 3079567
Manufacturer (Section G)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il 30795 67
IS 3079567
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key12307823
MDR Text Key266029587
Report Number3005075696-2021-00093
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup,Followup,Followup
Report Date 11/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberTPL0059
Device Catalogue NumberTPL0059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/09/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 08/11/2021 Patient Sequence Number: 1
-
-