Model Number IPN001112
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/14/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that system error 3 alarm occurred while on patient. As a result, the intra-aortic balloon pump (iabp) was swapped out. There was no report of patient complications, serious injury or death. Biomed also stated constant tone alarm and iabp shutting off occurred while testing in biomed. Fsa found system error 3/high baseline alarm after pumping for 2 minutes. Replacing the pump assembly fixed the high baseline alarm. Fsa could not replicate constant tone alarm and pump shutting off. Performed functional checklist, iabp performed to spec.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAC3 OPTIMUS IABP NA/EMEA
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kaylia pen
16 elizabeth drive
chelmsford, MA 01824
MDR Report Key12312924
MDR Text Key266205021
Report Number3010532612-2021-00218
Device Sequence Number1
Product Code DSP
UDI-Device Identifier30801902084966
UDI-Public30801902084966
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 07/14/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberIPN001112
Device Catalogue NumberIAP-0700
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/02/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-