It was reported that the machine could not generate the tidal volumes that were set on the ventilator.No adequate manual ventilation of the patient possible.Ambu bag was used to ventilate the patient.The patient was getting a mandibular fixation, and the surgery was stopped to do this.The patient desaturated to the 50's and it was decided to orally intubate the patient.Subsequently, the patient was transferred to the icu as a continued intermittently desaturation due to tube exchange was observed.
|
The device was checked in follow-up of the event by a dräger service engineer.The error condition could not be duplicated.Since it was stated that the device wouldn¿t have alarmed, especially the alarm functionality was tested in cooperation with the hospital¿s biomed.It was found working according to specifications; all provoked alarm conditions triggered the expected alarms.Evaluation of the log file made by the manufacturer confirms the on-site expertise in every detail.The records therein indicate that the device passed the system test in the morning of the doe without deviations.Just from the beginning of automatic ventilation in the concerned procedure the device alarmed for apnea, mv low and inspiratory tidal volume high indicating the presence of a significant circuit leak.This leakage was not removed and thus, it was finally leading to a fresh gas deficit; the corresponding fresh gas low or leakage alarm was posted.The device was set to standby mode for a few minutes then.The user initiated a new instance of the breathing system test which was passed.Most probably, the patient circuit has been replaced before this test as it was mentioned as a side note in the user¿s report.Afterwards the procedure was continued for more than 2 hours with stable and unremarkable ventilation towards the end.All in all, no indication for the potential presence of a device malfunction could be found.The reported symptom was most likely caused by a significant circuit leaks outside the device leading to a loss of pressure, volume and fresh gas.Dräger finally concludes that the device issued appropriate alarms to call the attention of the user.All alarms, possible causes as well as remedies are described in the instructions for use.
|
It was reported that the machine could not generate the tidal volumes that were set on the ventilator.No adequate manual ventilation of the patient possible.Ambu bag was used to ventilate the patient.The patient was getting a mandibular fixation, and the surgery was stopped to do this.The patient desaturated to the 50's and it was decided to orally intubate the patient.Subsequently, the patient was transferred to the icu as a continued intermittently desaturation due to tube exchange was observed.
|