• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTSINE TECHNOLOGIES LTD PACKAGE,500P,PP03,INTL EN,500-BAS-UK-10; AUTOMATED EXTERNAL DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTSINE TECHNOLOGIES LTD PACKAGE,500P,PP03,INTL EN,500-BAS-UK-10; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number SAM500P
Device Problems Defective Component (2292); Key or Button Unresponsive/not Working (4063)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/29/2021
Event Type  malfunction  
Manufacturer Narrative
Heartsine has requested return of the device for investigation.Upon completion, the conclusions will be submitted in a follow-up report.
 
Event Description
Inability to use the device shock button, may prevent or delay defibrillation which could result in adverse consequences.No patient involvement.
 
Manufacturer Narrative
Heartsine evaluated the customer's device and verified the event was logged in the device's memory; however, the reported issue could not be duplicated in testing, and the investigation was inconclusive as no fault was found on the returned device.The device was scrapped by heartsine and the customer was provided with a replacement device.
 
Event Description
The distributor contacted heartsine to report that their device was showing the "shock key stuck" error message.Inability to use the device shock button, may prevent or delay defibrillation which could result in adverse consequences.There was no patient involvement reported with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKAGE,500P,PP03,INTL EN,500-BAS-UK-10
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI  BT3 9ED
MDR Report Key12317827
MDR Text Key267974894
Report Number3004123209-2021-00206
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
PMA/PMN Number
P160008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 10/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberSAM500P
Device Catalogue Number500-BAS-UK-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2021
Date Manufacturer Received10/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-