|
Model Number 72203521 |
Device Problem
Flaked (1246)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/14/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Internal complaint reference case (b)(4).
|
|
Event Description
|
It was reported that during arthroscopy procedure, inside the patient, the 4.5mm full radius platinum series was leaving marks on patient's tissue.The procedure was finished with a smith and nephew backup device.No surgical delay.Patient injuries were not reported.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).H3, h6: the reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as pre-cautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A visual inspection revealed the device was returned outside of original packaging.The device was coated in debris at the distal tip.Signs of wear visible.No physical damage visible.A functional evaluation revealed that the device functioned as intended clockwise, counter clockwise and while oscillating.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
|
Search Alerts/Recalls
|
|
|