Model Number 500-56112 |
Device Problems
Device Alarm System (1012); Obstruction of Flow (2423)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/21/2021 |
Event Type
malfunction
|
Event Description
|
It was reported a drug occlusion occurred and required an additional procedure.Two ekosonic kit 106cm 12cm tz catheters were placed in the groin through an unspecified 6f sheath for the treatment of a bilateral pulmonary embolism case.After seven hours of therapy, the coolant pump was alarming on one of the catheters.Troubleshooting was attempted but was unsuccessful.While trying to fix the coolant occlusion issue, the catheter placed in the other pulmonary artery started alarming on the drug port.Again troubleshooting was attempted but was unsuccessful.The catheters were removed and the patient still had residual clotting so a second ekos therapy procedure had to be performed the following day.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: the ekos catheter was returned to boston scientific for analysis.The device was inspected for any damage or irregularities.The complaint was for drug infusion pump alerts and obstruction within device.There was blood noted in the drug luer.The device showed 13 of the 15 drug holes occluded with 2 holes clear.The device was run in a water bath for 15 minutes during which most of the drug holes opened and no alarms presented.Blood was visible in the drug luer but did not cause resistance when flushing during analysis.Foreign matter was seen embedded in the infusion catheter 10.5 cm from the distal end but was not blocking any drug holes.The ultrasonic core was able to traverse the length of the infusion catheter during analysis.Inspection of the remainder of the device revealed no damage or irregularities.Device analysis determined the condition of the returned device was not consistent with the reported information.The reported complaint of difficult to flush and obstruction within device were not confirmed.
|
|
Event Description
|
It was reported a drug occlusion occurred and required an additional procedure.Two ekosonic kit 106cm 12cm tz catheters were placed in the groin through an unspecified 6f sheath for the treatment of a bilateral pulmonary embolism case.After seven hours of therapy, the coolant pump was alarming on one of the catheters.Troubleshooting was attempted but was unsuccessful.While trying to fix the coolant occlusion issue, the catheter in the other pulmonary artery started alarming on the drug port.Again troubleshooting was attempted but was unsuccessful.The catheters were removed and the patient still had residual clot so a second ekos therapy procedure had to be performed the following day.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: the ekos catheter was returned to boston scientific for analysis.The device was inspected for any damage or irregularities.The complaint was for drug infusion pump alerts and obstruction within device.There was blood noted in the drug luer.The device showed 13 of the 15 drug holes occluded with 2 holes clear.The device was run in a water bath for 15 minutes during which most of the drug holes opened and no alarms presented.Blood was visible in the drug luer but did not cause resistance when flushing during analysis.Foreign matter was seen embedded in the infusion catheter 10.5 cm from the distal end but was not blocking any drug holes.The ultrasonic core was able to traverse the length of the infusion catheter during analysis.Inspection of the remainder of the device revealed no damage or irregularities.Device analysis determined the condition of the returned device was not consistent with the reported information.The reported complaint of difficult to flush and obstruction within device were not confirmed.
|
|
Event Description
|
It was reported a drug occlusion occurred and required an additional procedure.Two ekosonic kit 106cm 12cm tz catheters were placed in the groin through an unspecified 6f sheath for the treatment of a bilateral pulmonary embolism case.After seven hours of therapy, the coolant pump was alarming on one of the catheters.Troubleshooting was attempted but was unsuccessful.While trying to fix the coolant occlusion issue, the catheter in the other pulmonary artery started alarming on the drug port.Again troubleshooting was attempted but was unsuccessful.The catheters were removed and the patient still had residual clot so a second ekos therapy procedure had to be performed the following day.
|
|
Search Alerts/Recalls
|