Model Number 3186 |
Device Problems
Fracture (1260); High impedance (1291)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 07/27/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Date of event is estimated.
|
|
Event Description
|
Related manufacturer reference number: 3006705815-2021-03986.It was reported the patient experienced ineffective therapy.Diagnostics discovered multiple contacts with high impedance.X-rays showed a possible fracture.In turn, surgical intervention may take place at a later date to address the issue.
|
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
Additional information received confirmed that both leads fractured.The patient underwent surgical intervention wherein both existing leads were explanted and replaced to address the issue.
|
|