Catalog Number 046W1AN00641
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/19/2021
Event Type  malfunction  
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event Description
It was reported that the tip of a final driver broke off intra-operatively.There was a delay without patient impacts while the tip was retrieved.
 
Manufacturer Narrative
Corrections in d4: udi number, d9, and h3.Additional information in h4 and h6: component, investigation type, findings, and conclusions.Device evaluation: visual inspection revealed the tip of the device has fractured.Potential cause root cause was unable to be determined.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.It could also be attributed to off-axis forces exerted on the device during use.Dhr review per dhr review, the part was likely conforming when it left zimmer biomet control.Device use this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event Description
It was reported that the tip of a final driver broke off intra-operatively.There was a delay without patient impacts while the tip was retrieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FINAL SCREWDRIVER SHAFT II
Type of Device
INSTINCT JAVA SYSTEM
Manufacturer (Section D)
ZIMMER SPINE
23 parvis des chartrons
cite mondiale
bordeaux, cedex 33080
FR  33080
Manufacturer (Section G)
ZIMMER SPINE
23 parvis des chartrons
cite mondiale
bordeaux, cedex 33080
FR   33080
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key12325118
MDR Text Key266675418
Report Number3003853072-2021-00090
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K111301
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number046W1AN00641
Device Lot NumberA2190909A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Weight61 KG
-
-