Additional narrative: udi: (b)(4).
Investigation summary: the complaint device was received at the service center and evaluated.
It was reported that there is a crack inside the device.
Per service reports, this complaint can be confirmed.
During the service evaluation the following defects were identified: cable in good condition.
Keyboard resistance value out of tolerance limits.
Corroded motor.
The motor and keyboard were replaced to resolve the issues.
After repair, the device was found to be working according to the specifications.
The faulty parts was identified as the root cause for the device failure during the service evaluation.
Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
At this point in time, no corrective action is required, and no further action is warranted.
Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|