DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/24/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device (10): a getinge field service engineer (fse) evaluated the iabp and was unable to reproduce the reported issue.However, the fse found during the pressure regulator test that the pressure was slow coming up to pressure of 390 mmhg and during the compressor output test the compressor¿s temperature felt hot to the touch.The fse replaced the compressor and functional tested it without any further issues.All calibration, functional and safety checks to meet factory specifications was performed.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted when our investigation is completed.
|
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) had a system over temperature alarm.The iabp was switched to another without incident.The iabp had been tagged as inoperable until service is completed.No patient harm, serious injury or adverse event was reported.
|
|