It was reported that the site experienced a longer startup time for the centrimag console.The system was rebooted, after that, the console worked properly.No patient was involved, this occurred when the healthcare professional did the initial boot up of the system.
|
Section h6 health effect - impact code: "2199 - no health consequences or impact" corrected to "2645 - no patient involvement".Manufacturer's investigation conclusion: the reported event of the centrimag console having a longer startup time was not confirmed.The centrimag console (serial number (b)(6)) was not returned for analysis, and no log files were associated with the reported event.The issue was stated to have resolved upon rebooting the system, and, per additional information, the console would be kept in the hospital.The root cause of the reported event was unable to be determined through this analysis.Review of the device history record for the centrimag console showed the device was manufactured in accordance with manufacturing and quality assurance specifications.The 2nd generation centrimag system operating manual section 3 entitled ¿about the 2nd generation centrimag primary console" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section 8 entitled "emergency/troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." no further information was provided.The manufacturer is closing the file on this event.
|