Model Number CV-180
Device Problems No Display/Image (1183); Electrical /Electronic Property Problem (1198); Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2021
Event Type  malfunction  
Manufacturer Narrative
The subject device was not returned to omsc for evaluation but was returned to (b)(4). (b)(4) checked the subject device and found that the subject device did not start up and the endoscopic image of the subject device was not displayed as reported due to the failure of the printed circuit board, and also found that the front panel led lit up continuously as reported. The exact cause has been under investigation. Therefore, the exact cause of the reported event could not be conclusively determined at this time. If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp. (omsc) was informed from the user facility that during routine inspection by the user, it was found that the endoscopic image of the subject device was not displayed on the monitor. The field service engineer of olympus medical systems (b)(4) checked the subject device on-site and found that the reported phenomenon was duplicated, which might be caused by the malfunction of the subject device, and also found that the front panel led lit up continuously. There was no report of patient injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameEVIS EXERA II VIDEO SYSTEM CENTER
Type of DeviceVIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12336181
MDR Text Key267886954
Report Number8010047-2021-10408
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 10/28/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberCV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received09/30/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2011
Is the Device Single Use? No Answer Provided
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-