BOSTON SCIENTIFIC CORPORATION FILTERWIRE EZ; TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE Back to Search Results
Model Number 44150
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/30/2021
Event Type  malfunction  
Event Description
It was reported that the delivery sheath was separated upon removal.The 70% stenosed target lesion was located in the moderately tortuous and mildly calcified common carotid artery.A filter wire ez has been selected for use.During deployment, the filter wire has crossed the stenosed area.However, resistance was felt when the sheath touched the y-connector that was attached into the guiding catheter.When the device was pulled strongly, the delivery sheath got separated.The separated sheath remained on the filter wire so the subsequent procedure could not be performed.The filter wire was then removed together with the separated delivery sheath.The procedure was completed with another of same device.No complications reported and patient was in good health soon after the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FILTERWIRE EZ
Type of Device
TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12336909
MDR Text Key267080648
Report Number2134265-2021-10291
Device Sequence Number1
Product Code NTE
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 08/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/01/2023
Device Model Number44150
Device Catalogue Number44150
Device Lot Number0026892036
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/30/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
-
-