OLYMPUS MEDICAL SYSTEMS CORP. ULTRASONIC PROBE Back to Search Results
Model Number UM-S20-17S
Device Problems No Display/Image (1183); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/21/2021
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to omsc for evaluation.The ultrasonic image was hardly drawn.There was a hole on the sheath at the tip, and the ultrasonic medium leaked out.There was a dent on the sheath at the tip.The inside color was changed.The internal blade could be driven normally.The manufacturing record was reviewed and found no irregularities.Omsc presumed that the ultrasonic image was hardly drawn because the ultrasonic medium inside the sheath leaked out of the damaged part of the sheath.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the preparation for use, the ultrasonic image was not drawn although the vibrator of the probe was rotating.The subject device was used in combination with eu-me1.There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12340584
MDR Text Key267274468
Report Number8010047-2021-10465
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K982323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberUM-S20-17S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/27/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/23/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-