• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) OCTRODE LEAD KIT, 60CM LENGTH; PERCUTANEOUS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) OCTRODE LEAD KIT, 60CM LENGTH; PERCUTANEOUS LEAD Back to Search Results
Model Number 3186
Device Problems Fracture (1260); High impedance (1291)
Patient Problem Inadequate Pain Relief (2388)
Event Date 07/29/2021
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.
 
Event Description
Related manufacturer reference number: 3006705815-2021-03893.It was reported that stimulation on the leads was reducing on its own.Both leads displayed high impedance.X-rays revealed potential fracture in one of the lead.Surgical intervention may take place to address the issue.
 
Manufacturer Narrative
During processing of this complaint, attempts were made to obtain weight but not received.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Additional information received indicated the patient underwent surgical intervention wherein both of the leads were explanted and replaced to address the issue.Therapy was restored postoperatively.
 
Manufacturer Narrative
The reported event of high impedances was confirmed.Both leads were returned complete.Microscopic inspection identified a kink in the lead body where all of the internal wires were broken followed by a cam impression mark.The fracture wires are consistent with overstress condition that leads may have been subjected to while in viv.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
PERCUTANEOUS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
MDR Report Key12341633
MDR Text Key267238573
Report Number3006705815-2021-03894
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067017246
UDI-Public05415067017246
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup,Followup
Report Date 10/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/11/2022
Device Model Number3186
Device Catalogue Number3186
Device Lot NumberA000103987
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PERCUTANEOUS LEAD; PERCUTANEOUS LEAD
Patient Outcome(s) Other;
-
-