• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION JETSTREAM XC; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION JETSTREAM XC; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 45007
Device Problems Difficult to Advance (2920); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/28/2021
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by manufacturer: the device returned to boston scientific consisted of a jetstream xc-2.4 atherectomy catheter.The device was visually and microscopically examined for any damage.Visual examination showed multiple catheter shaft damage in the form of buckling.The device showed a burst infusion sheath 21.8cm from the strain relief.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
 
Event Description
It was reported that infusion line material deformation occurred.A jetstream xc catheter was prepared, and once inside the patient's body, the catheter was hard to advance.They noticed the outside of the catheter started to buckle.Eventually the priming liquid was squirting out of the catheter.A new device was used to complete the case with no complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM XC
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12342349
MDR Text Key267265966
Report Number2134265-2021-10522
Device Sequence Number1
Product Code MCW
UDI-Device Identifier08714729889922
UDI-Public08714729889922
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/04/2023
Device Model Number45007
Device Catalogue Number45007
Device Lot Number0027248337
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/28/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age49 YR
-
-