Model Number TABLETOP
Device Problem Component Missing (2306)
Patient Problem Retinal Injury (2048)
Event Date 07/15/2021
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that an ophthalmic operating console's laser filter was missing during a surgery.The surgeon had left macular laser injury.The retina specialist examined the eye and optical coherence tomography (oct) and angiogram were performed which showed retinal pigment epithelium (rpe) damage.The acute symptoms have been resolved but there is a possibility of long-term choroidal neovascularization (cnv).
 
Manufacturer Narrative
The customer did not request service for the system.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to user handling.Data will continue to be monitored for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key12342808
MDR Text Key267280921
Report Number2028159-2021-00920
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657511501
UDI-Public00380657511501
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Was Device Available for Evaluation? No
Date Manufacturer Received09/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PUREPOINT LASER; PUREPOINT LASER
Patient Outcome(s) Other;
-
-