Model Number 3186
Device Problems Fracture (1260); Migration (4003)
Patient Problem Inadequate Pain Relief (2388)
Event Date 04/07/2021
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated. Based on the information provided a device problem was not identified. As a result, a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
 
Event Description
Related manufacturer reference numbers: 3006705815-2021-04108, 3006705815-2021-04110. It was reported that the patient lost weight, the ipg was unable to be set to mri mode, and patient experienced ineffective therapy. Diagnostics revealed low impedance and high impedance on the leads. As a result, surgery occurred to address the issues. During the surgery, the leads were found to be twisted and fractured and the ipg may have twisted. During the surgery, the system was explanted and replaced to address the issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameOCTRODE LEAD KIT, 60CM LENGTH
Type of DevicePERCUTANEOUS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key12346476
MDR Text Key267414233
Report Number3006705815-2021-04109
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation
Type of Report Initial
Report Date 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/20/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date09/11/2021
Device Model Number3186
Device Catalogue Number3186
Device Lot NumberA000086362
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/29/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 08/20/2021 Patient Sequence Number: 1
Treatment
PERCUTANEOUS LEAD; SCS IPG
-
-