ARTHROCARE CORP. DISPOSABLE KIT HIP FOR 1.8MM Q-FIX IMP; TROUSERS, ANTI-SHOCK Back to Search Results
Catalog Number 25-1811
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Event Description
It was reported that during an arthroscopy, the drill from the 1.8 q-fix suture anchor disposable kit broke off.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).H10 h3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE KIT HIP FOR 1.8MM Q-FIX IMP
Type of Device
TROUSERS, ANTI-SHOCK
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key12351074
MDR Text Key267617122
Report Number3006524618-2021-00857
Device Sequence Number1
Product Code LHX
UDI-Device Identifier00885556724620
UDI-Public885556724620
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 09/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/20/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/22/2024
Device Catalogue Number25-1811
Device Lot Number2067044
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-