MAUDE Adverse Event Report: DATEX-OHMEDA, INC. PANDA IRES WARMER; WARMER, INFANT RADIANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. PANDA IRES WARMER; WARMER, INFANT RADIANT Back to Search Results
Device Problem Material Integrity Problem (2978)
Patient Problem Insufficient Information (4580)
Event Date 07/27/2021
Event Type  malfunction  
Manufacturer Narrative
No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Ge healthcare (gehc) reported a field modification for this issue per 21 cfr 806 on 20 march 2019.The fda recall number is z-1846-2019 & z-1847-2019.Customers were sent a letter explaining the issue and requesting the customer to inspect the warmer bedside panel latch areas.Replacement of broken bedside panels will be provided by gehc.A set of warning labels will be supplied for application to the bedside panels.These labels will warn the user to not use the bedside panels for maneuvering the warmer and indicate the correct method of maneuvering the warmer.An addendum to the operation and maintenance manual will also be provided emphasizing the need to check and ensure that the bedside panels and latches are not cracked, broken, or damaged before every patient use.The addendum will also contain instructions to increase detectability of broken or cracked bedside panels.On june 14, 2019 a letter was sent to customers providing an additional addendum, labels, and wall poster.
 
Event Description
As a result of an inspection that was completed as part of correction and removal z-1846-2019 & z-1847-2019, this system was identified as having the broken latch and/or cracked wall malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PANDA IRES WARMER
Type of Device
WARMER, INFANT RADIANT
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key12354057
MDR Text Key268922027
Report Number2112667-2021-01940
Device Sequence Number1
Product Code FMT
Combination Product (y/n)N
PMA/PMN Number
K122267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Remedial Action Recall
Type of Report Initial
Report Date 08/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/27/2021
Date Device Manufactured02/04/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberSEEH10
Patient Sequence Number1
-
-