• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH VIPER TI SAI POLY 9X100MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH VIPER TI SAI POLY 9X100MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD Back to Search Results
Model Number 179704999
Device Problem Break (1069)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).Additional product code: kwp;kwq;mnh;mni.Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Without a lot number, the device history records review could not be completed as no product was received.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date, the patient had a surgery for posterior spinal fusion in (b)(6) of 2020.At a follow up appointment in (b)(6) of 2020, xray images showed that the set screw had loosened from the sai screws bilaterally and the rod had loosened due to head splay.The s1 and l5 screws had fractured at the neck of the screw just below the tulip.It is unknown if fragments were generated.The patient did not achieve a fusion below l5 due to this hardware failure.Procedure outcome is unknown.This complaint involves five (5) devices.This report is for (1) viper ti sai poly 9x100mm.This report is 3 of 5 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER TI SAI POLY 9X100MM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kara ditty-bovard
chemin-blanc 38
le locle, MA 
SZ  
6103142063
MDR Report Key12354169
MDR Text Key267681393
Report Number1526439-2021-01748
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034265889
UDI-Public10705034265889
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160904
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number179704999
Device Catalogue Number179704999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/24/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
5.5 TI CORT FIX 7X45MM; 5.5 TI CORT FIX 7X45MM; 5.5 TI CORT FIX 7X45MM; VIPER TI SAI POLY 9X90MM
Patient Outcome(s) Required Intervention;
-
-