It was reported during treatment with two (2) units of a prismaflex st150 set and a prismaflex control unit , a decent sized hole was observed in the bag causing it to leak and empty quickly.
An air in blood alarm was generated, however, the bag empty alarm was not triggered.
The treatment was terminated without the extracorporeal (ec) blood being returned to the patient.
The treatment was restarted with a new set, and after six (6) hours of treatment, the pre blood pump bag had run dry and again an air in blood alarm was triggered.
The treatment was once again terminated without returning the ec blood to the patient.
There was no patient injury or medical intervention associated with this event.
No additional information is available.
|