Model Number 3771 |
Device Problem
Temperature Problem (3022)
|
Patient Problem
Burning Sensation (2146)
|
Event Date 08/18/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Date of event is estimated.
|
|
Event Description
|
It was reported that the patient was experiencing a burning/shocking sensation at the ipg site when charging and when therapy is on.Imaging was done which showed excess lead wires being coiled up on top of the ipg between the ipg and skin.In turn, surgical intervention may take place to address the issue.
|
|
Event Description
|
Additional information revealed that the patient underwent surgical intervention wherein the ipg was explanted and replaced, resolving the issue.
|
|
Manufacturer Narrative
|
Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
|
|