Model Number 3771
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Date 08/18/2021
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.
 
Event Description
It was reported that the patient was experiencing a burning/shocking sensation at the ipg site when charging and when therapy is on.Imaging was done which showed excess lead wires being coiled up on top of the ipg between the ipg and skin.In turn, surgical intervention may take place to address the issue.
 
Event Description
Additional information revealed that the patient underwent surgical intervention wherein the ipg was explanted and replaced, resolving the issue.
 
Manufacturer Narrative
Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGE MRI IPG
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
MDR Report Key12356033
MDR Text Key267748325
Report Number3006705815-2021-04151
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067017086
UDI-Public05415067017086
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/05/2018
Device Model Number3771
Device Catalogue Number3771
Device Lot NumberA000030002
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight59
-
-