MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP-JAPAN
Device Problems Display or Visual Feedback Problem (1184); Operating System Becomes Nonfunctional (2996)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/29/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that during a cataract vitrectomy surgery, an ophthalmic operating console screen turned white and the console stopped working.The console was restarted, but the issue was not resolved.An alternate console was obtained in order to complete the surgery.There was no patient harm.
 
Manufacturer Narrative
The company service representative examined the system and replicated the reported event.The company service representative found an issue with the memory of the host module.The company service representative performed cleaning and lubrication of the module to resolve the issue.The system was then tested and found it to perform as intended.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to component wear of the host module.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key12362321
MDR Text Key267960990
Report Number2028159-2021-00937
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 10/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP-JAPAN
Device Catalogue Number8065751817
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-