Model Number 16-02-82
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2021
Event Type  Malfunction  
Manufacturer Narrative

There was no patient involvement. Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). A livanova field service representative was dispatched to the facility to investigate the device and found out that stir motor bearing to be seized. Replaced the stir motor, replaced the cardioplegia pump on the cold side, tested the machine and it was back to operating properly. Calcification on unit not to level requiring de-calcification. The device was warming only between 31 and 32 degrees. The temperature sensors were replaced and the heaters placed on order to complete the repair. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.

 
Event Description

Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error code 7 immediately after turn on. There was no report of patient injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key12365403
MDR Text Key268063298
Report Number9611109-2021-00458
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation
Type of Report Initial
Report Date 08/25/2021
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received08/25/2021
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number16-02-82
Was Device Available For Evaluation? No Answer Provided
Is The Reporter A Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/26/2021
Was Device Evaluated By Manufacturer? No Answer Provided
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-