MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 97714
Device Problems Intermittent Continuity (1121); Failure to Deliver Energy (1211); Therapy Delivered to Incorrect Body Area (1508); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Electric Shock (2554)
Event Date 08/24/2021
Event Type  Malfunction  
Manufacturer Narrative

If information is provided in the future, a supplemental report will be issued.

 
Event Description

Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use. It was reported that for a month the patient felt the stimulation unexpectedly stopped and had to use his remote control to turn it on; this happened twice in august. During the same time the patient describe having electricity sensations throughout his body several times a day. This did not occur during a particular activity or location. There were no external contributing factors found but a contributing factor was ankylosing spondylitis. Troubleshooting was performed: measured impedances of system (measurements within normal range), asked the patient if something happened before the feeling of electricity (fall, car accident, use of machine that creates magnetic field). The patient was asked to turn the ins off for a couple days and see if he still has the feeling of electricity. The issue was not resolved at the time of this report. The patient was awaiting neurological examinations to see what could be the origin of the electricity feeling. No surgical intervention occurred. The patient was alive with no injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSURESCAN
Type of DeviceSTIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key12370696
MDR Text Key268244595
Report Number3004209178-2021-12949
Device Sequence Number1
Product Code LGW
Combination Product (Y/N)N
Reporter Country CodeFR
PMA/PMN NumberP840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,FOREIG
Reporter Occupation
Type of Report Initial
Report Date 08/26/2021
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received08/26/2021
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device EXPIRATION Date10/28/2016
Device MODEL Number97714
Device Catalogue Number97714
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/24/2021
Was Device Evaluated By Manufacturer? No Answer Provided
Date Device Manufactured12/10/2015
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-