• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC UNK - PLATES: FNS PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC UNK - PLATES: FNS PLATE, FIXATION, BONE Back to Search Results
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/29/2021
Event Type  Malfunction  
Manufacturer Narrative

This report is for an unk - plates: fns/unknown lot. Part and lot numbers are unknown; udi number is unknown. Complainant part is not expected to be returned for manufacturer review/investigation. Reporter is a j&j sales representative. Without a lot number the device history records review could not be completed. Product was not returned. Based on the information available, it has been determined that no corrective and/or preventative action is proposed. This complaint will be accounted for and monitored via post market surveillance activities. If additional information is made available, the investigation will be updated as applicable. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.

 
Event Description

It was reported that on (b)(6) 2021 the patient underwent removal surgery due to patient collapsed into theirs in the device cut out. During the removal, one of the locking screws is cold-welded to the plate so we have to cut the plate with the metal cutting tool. It is unknown if there was a surgical delay. Procedure was completed successfully without any surgical delay. Concomitant devices reported: unknown fns plate (part# unknown, lot# unknown, quantity 1). This report is for 1 unk - plates: fns. This is report 2 of 2 for complaint (b)(4).

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNK - PLATES: FNS
Type of DevicePLATE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key12375139
MDR Text Key268457965
Report Number2939274-2021-04966
Device Sequence Number1
Product Code HRS
Combination Product (Y/N)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Type of Report Initial,Followup
Report Date 07/29/2021
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received08/26/2021
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/09/2021
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? No Answer Provided
Is this a Reprocessed and Reused Single-Use Device?
Type of Device Usage Unkown

-
-