Manufacturer Narrative
There was no known patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is (b)(4).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t device was found to be contaminated with mycobacterium gordonae.There is no known patient involvement.
 
Manufacturer Narrative
H.10: through follow-up communication under recent similar complaints from the same hospital livanova learned that the device is cleaned regularly according to the instruction for use and positioned outside the operating theatre during use.However, livanova learned that hydrogen peroxide h2o2 level is not checked daily as prescribed by the ifu and that is not monitored during period of non use even if the device is stored full of water.This is not in accordance with device instruction for use and this deviation may have led/contributed to the reported contamination.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: originally it was communicated to livanova that the device was contaminated with m.Gordonae.Through follow-up communication livanova retrieved the laboratory report.Analysis of the provided report show that the device was contaminated with m.Gordonae and paragordonae.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key12376263
MDR Text Key268451646
Report Number9611109-2021-00472
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeBE
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 08/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/19/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015
Patient Sequence Number1
-
-