Model Number SC-1200 |
Device Problems
Premature Discharge of Battery (1057); Charging Problem (2892)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/04/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Date of event: approximated based on the date the manufacturer became aware of the event.
|
|
Event Description
|
It was reported that the patient had difficulty charging the ipg.The physician suspected premature battery depletion due to electrocautery damaged.The patient underwent an ipg replacement procedure and was doing well post-operatively.Per the physicians implant manual, electrocautery is a known source of high voltage transient signal and warns against the use of electrocautery.
|
|
Event Description
|
It was reported that the patient had difficulty charging the ipg.The physician suspected premature battery depletion due to electrocautery damaged.The patient underwent an ipg replacement procedure and was doing well post-operatively.Per the physicians implant manual, electrocautery is a known source of high voltage transient signal and warns against the use of electrocautery.
|
|
Manufacturer Narrative
|
Sc-1200 sn: (b)(6).The returned ipg was analyzed, passed all tests performed, and exhibited normal device characteristics.
|
|