• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION PRECISION MONTAGE MRI; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION PRECISION MONTAGE MRI; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-1200
Device Problems Premature Discharge of Battery (1057); Charging Problem (2892)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/04/2021
Event Type  Injury  
Manufacturer Narrative
Date of event: approximated based on the date the manufacturer became aware of the event.
 
Event Description
It was reported that the patient had difficulty charging the ipg.The physician suspected premature battery depletion due to electrocautery damaged.The patient underwent an ipg replacement procedure and was doing well post-operatively.Per the physicians implant manual, electrocautery is a known source of high voltage transient signal and warns against the use of electrocautery.
 
Event Description
It was reported that the patient had difficulty charging the ipg.The physician suspected premature battery depletion due to electrocautery damaged.The patient underwent an ipg replacement procedure and was doing well post-operatively.Per the physicians implant manual, electrocautery is a known source of high voltage transient signal and warns against the use of electrocautery.
 
Manufacturer Narrative
Sc-1200 sn: (b)(6).The returned ipg was analyzed, passed all tests performed, and exhibited normal device characteristics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION MONTAGE MRI
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
MDR Report Key12376564
MDR Text Key268427219
Report Number3006630150-2021-04793
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729905943
UDI-Public08714729905943
Combination Product (y/n)N
PMA/PMN Number
P030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/19/2021
Device Model NumberSC-1200
Device Catalogue NumberSC-1200
Device Lot Number359750
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2021
Date Manufacturer Received09/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age58 YR
-
-