Investigation of returned product: typical values for freshly produced dental dam are tensile greater than or equal to 3,500 lbf/in² and ultimate elongation greater than or equal to 700% with reference to typical modulus at 100% elongation at 120 lbf/in².Properties will deteriorate as the dam ages.The mean value for tensile is 4,414 lbf/in² which is within specification for newly produced dental dam as well as ultimate elongation at 788% with 110 lbf/in² as the modulus at 100% lbf/in².From the data produced, it is determined that the dental dam is as expected and no abnormal results were found.The fit, form and function of the dam is not compromised.A review of past complaints was performed regarding tears for this sheeting lot from may 2019 until present.This is the first complaint recorded for this lot/part; therefore, no recurrence was identified.
|