MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL (SINGAPORE) SYRINGE 1ML LS TUBERCULIN; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL (SINGAPORE) SYRINGE 1ML LS TUBERCULIN; PISTON SYRINGE Back to Search Results
Catalog Number 302100
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/29/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(6).A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported syringe 1ml ls tuberculin had scale marking issues.The following information was provided by the initial reporter, translated from (b)(6): "some syringe memory is double printed.I don't think this can be measured accurately.".
 
Manufacturer Narrative
H.6.Investigation: four photos and forty-four (44) samples were received by our quality team for evaluation.From the photos, illegible scale line printing (printing overlap) was observed.Twenty-five of the returned samples did not belong to tuas manufacturing.The remaining nineteen samples were subjected to visual inspection and illegible scale line printing (printing overlap) was observed.A review of the internal manufacturing device records and raw material history files for the reported lot numbers was performed and no recorded quality problems or rejections to this incident were found.The probable root cause of this nonconformance could be that the mandrel bearing was worn-out.This causes the mandrels rotation to not be smooth and create a jerky motion.The jerky mandrels with play cause the scale line on the printing pad being transferred twice when the mandrels move left and right, thereby overriding the scale on the barrel surface.
 
Event Description
It was reported syringe 1ml ls tuberculin had scale marking issues.The following information was provided by the initial reporter, translated from japanese: "some syringe memory is double printed.I don't think this can be measured accurately.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 1ML LS TUBERCULIN
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
MDR Report Key12378666
MDR Text Key268510224
Report Number8041187-2021-00777
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 09/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number302100
Device Lot Number0157814
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2021
Date Manufacturer Received09/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-