• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION UNKNOWN OPTEASE VENA CAVA FILTER FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION UNKNOWN OPTEASE VENA CAVA FILTER FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number 466FXXXX
Device Problems Unintended Movement (3026); Migration (4003)
Patient Problem Perforation (2001)
Event Date 08/04/2021
Event Type  Injury  
Manufacturer Narrative

As reported, the patient underwent placement of an optease retrievable vena cava filter. The indication for the filter placement was not reported. At some point after the filter implantation, the patient became aware that the filter tilted and migrated and was associated with perforation. The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed. The optease retrievable vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the inferior vena cava (ivc) for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated. The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart. Without images or procedural films for review, the reported filter tilt, migration and perforation events could not be confirmed and the exact cause could not be determined. Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuousness. Additionally, the timing and mechanism of the filter tilt is unknown. Ivc filter migration is a known potential adverse event associated with all ivc filter implants and is listed in the instruction for use (ifu) as such. Possible causes for filter migration include mega cava, wire entrapment during central venous catheter placement, ¿sail¿ effect (cranial migration) of large clot burden within the filter, mechanical device failure, and operator error. Physiologic causes of migration may result from temporary dysmorphism of the ivc including bending, coughing or valsalva maneuvers resulting in dislodgment of the filter. Some studies suggest that strenuous physical activity and increased intra-abdominal pressure can lead to migration of ivc filters. It is unknown if the tilt contributed to the reported perforation. A review of the ifu notes vessel damage such as intimal tears and perforation as procedural complications related to ivc filters. Perforation from removable filters is relatively common, and directly related to how long the filter has been in place. Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation. Clinical factors that may have influenced the event include the patient¿s pre-existing co-morbidities, pharmacological issues and lesion characteristics. Given the limited information available for review, there is nothing to suggest that a malfunction in the design and manufacturing process of the device; therefore, no corrective action will be taken. Should additional information become available, the file will be updated accordingly.

 
Event Description

As reported in the legal brief, a patient underwent placement of an optease vena cava filter. The filter subsequently malfunctioned and caused injury and damage to the patient, including, but not limited to physical and emotional damages from tilting of the filter, migration, perforation and the resultant symptoms. As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages and required medical care and treatment. As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, pain, suffering, loss of enjoyment of life, distress and other damages.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN OPTEASE VENA CAVA FILTER
Type of DeviceFILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60th avenue
miami lakes FL 33014 2802
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60th ave
miami lakes FL 33014 2802
Manufacturer Contact
karla castro
14201 nw 60th ave
miami lakes, FL 33014-2802
7863138372
MDR Report Key12378691
MDR Text Key268510339
Report Number1016427-2021-05347
Device Sequence Number1
Product Code DTK
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK034050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type OTHER
Reporter Occupation
Type of Report Initial
Report Date 08/27/2021
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received08/27/2021
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue Number466FXXXX
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/06/2021
Was Device Evaluated By Manufacturer? No Answer Provided
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 08/27/2021 Patient Sequence Number: 1
-
-