Model Number 0998-00-3023-53 - CS300 ENGLISH |
Device Problem
Charging Problem (2892)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/03/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device: a getinge field service engineer (fse) was dispatched to evaluate the iabp and could not reproduce the reported issue.The fse observed that the unit was working on battery properly but continued showing low battery indication on display.The fse determined that the problem was with the power supply of the unit.To fix the issue, the fse replaced the power supply charger and performed all functional and safety checks to meet factory specifications and checked the performance of the unit with trainer and balloon connected on battery as well as mains.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of the investigation.
|
|
Event Description
|
It was reported that during a routine check performed by the customer, the cs300 showing a low battery indicator when switched over to battery.There was no patient involvement, thus no adverse event reported.
|
|