Model Number 0998-00-3023-53 - CS300 ENGLISH
Device Problem Charging Problem (2892)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/03/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: a getinge field service engineer (fse) was dispatched to evaluate the iabp and could not reproduce the reported issue.The fse observed that the unit was working on battery properly but continued showing low battery indication on display.The fse determined that the problem was with the power supply of the unit.To fix the issue, the fse replaced the power supply charger and performed all functional and safety checks to meet factory specifications and checked the performance of the unit with trainer and balloon connected on battery as well as mains.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
It was reported that during a routine check performed by the customer, the cs300 showing a low battery indicator when switched over to battery.There was no patient involvement, thus no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
dorota wolpiuk
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12380458
MDR Text Key268779306
Report Number2249723-2021-01941
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 08/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53 - CS300 ENGLISH
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/03/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-