Model Number 2544-00-522 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/20/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that the one of the pin holes on the cut block has been damaged on the inside and thus a pin cannot go cleanly though it, rendering it useless.It was also reported that it was not used on a patient.No surgical delay.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: examination of the returned cutting block identified the presence of a burr inside one of the eight pin holes.The noted damage is consistent with device wear out due to normal intended use and the investigation did not establish a need for corrective action.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|