• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FLOWONIX MEDICAL, INC. INTRATHECAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FLOWONIX MEDICAL, INC. INTRATHECAL CATHETER Back to Search Results
Model Number 11823
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/05/2021
Event Type  Injury  
Manufacturer Narrative
A review of the device history record, which includes verification of all steps in the manufacturing of the us catheter kit, verification of all final testing performed by/on the us catheter kit, and packaging for subject us catheter kit was performed.The review did not identify any non-conformances, issues or capas associated with us catheter kit function.Device was not returned for additional evaluation and investigation.It was reported that the cause of the fracture was unknown.Per the instructions for use of the device, catheter fracture is a known possible risk of use of the device.Internal complaint number: (b)(4).
 
Event Description
Sales representative reported that a catheter fracture occurred due to a catheter fracture.No patient effects were reported and the reason for the fracture is unknown.It was reported that the surgeon was able to use a catheter revision kit to connect the catheter without cutting any of the catheter and simply spliced it together.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRATHECAL CATHETER
Type of Device
INTRATHECAL CATHETER
Manufacturer (Section D)
FLOWONIX MEDICAL, INC.
500 international drive
suite 200
mount olive NJ 07828
Manufacturer (Section G)
FLOWONIX MEDICAL, INC.
120 forbes blvd
suite 170
mansfield MA 02048
Manufacturer Contact
james bennett
500 international drive
suite 200
mount olive, NJ 07828
9734269229
MDR Report Key12384416
MDR Text Key268689242
Report Number3010079947-2021-00250
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00810335020099
UDI-Public(01)00810335020099(17)220909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/09/2022
Device Model Number11823
Device Catalogue Number11823
Device Lot Number26908
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/05/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/09/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-