BAXTER HEALTHCARE CORPORATION 2000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938740
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/05/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 2000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag leaked from the back through a pinhole.The leak was discovered at the pharmacy during set up.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4, h6 and h10.H4: the lot was manufactured from april 15, 2021 - april 16, 2021.H10: the device was received for evaluation.Unaided visual inspection was performed which observed a tear in the top back side of the bag.A functional testing was performed which revealed a leak only at the top back side of the bag.Additional magnified inspection was performed which verified a tear/hole where the leak occurred.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key12386038
MDR Text Key268909542
Report Number1416980-2021-05312
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477244
UDI-Public(01)00085412477244
Combination Product (y/n)Y
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/31/2024
Device Catalogue NumberH938740
Device Lot Number60291028
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2021
Date Manufacturer Received08/31/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-