• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-55
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: a getinge field service engineer (fse) was dispatched to investigate. The fse evaluated the iabp unit and determined that the power supply unit is defective. At request of the customer's biomedical engineer, the power supply was removed from a cs100 iabp unit and plugged into the faulty cs300 iabp unit. It was observed that the cs300 was then in working condition. Subsequently, the stm completed all safety, functionality, and calibration checks and all tests passed to factory specifications. The iabp unit was cleared for clinical use and released to the customer. A supplemental report will be submitted upon completion of our investigation. The full name of the event site was shortened due to field character limit; the full name is (b)(6).
 
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) was not working. No additional details were provided. It is unknown the circumstances under which the event occurred. It is also unknown if there was a patient involvement. However, there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12388395
MDR Text Key268977642
Report Number2249723-2021-01954
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/13/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-00-3023-55
Device Catalogue Number0998-00-3023-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/28/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse

-
-